Turbo tsh5
Home Up Calcitonin IRMA test Anti-hTG (human thyroglobulin) RIA test TG (thyroglobulin) IRMA test Anti-TPO (thyroid peroxidase) RIA test Turbo tsh 1 Turbo tsh 2 Turbo tsh3 Turbo tsh4 Turbo tsh5 TSH ELISA test Free T4 (thyroxine) RIA test Free T4 Blue T4 (thyroxine) RIA test T4 (thyroxine) RIA test 1 T4 (thyroxine) RIA test 2 T4 (thyroxine) RIA test 3 Free T3 (triiodothyronine) RIA test T3 (triiodothyronine) RIA test T3 (triiodothyronine) RIA test 1 T3 (triiodothyronine) RIA test 2 T3 (triiodothyronine) RIA test 3 Progesterone RIA test Testosterone RIA test Cortisol RIA test DHEA-SO4 (dehydroepiandrosterone sulphate) RIA test hGH (human growth hormone) IRMA test   a hGH (human growth hormone) IRMA test 1  b hLH (luteinizing hormone) IRMA test hLH (luteinizing hormone) IRMA test 1 h FSH (follicle stimulating hormone) IRMA test h FSH (follicle stimulating hormone) IRMA test 1 h Prolactin IRMA test h Prolactin IRMA test 1 Insulin RIA test C-peptide IRMA test SHBG (sex hormone binding globulin) IRMA test AFP (alpha fetoprotein) IRMA test AFP (alpha fetoprotein) IRMA test 1 AFP (alpha fetoprotein) ELISA test 2 ß hCG (beta human chorionic gonadotropin) IRMA test ß hCG (beta human chorionic gonadotropin) IRMA test 1 ß hCG (beta human chorionic gonadotropin) ELISA test 2 hCG (human chorionic gonadotropin) free ß-subunit test 3 h CEA (carcinoembryonic antigen) IRMA test PSA (prostate specific antigen) IRMA test PSA (prostate specific antigen) IRMA test 1 Free PSA (prostate specific antigen) IRMA test PSA (prostate specific antigen) ELISA test Prostaglandin D2 RIA test 8-iso-Prostaglandin F2a RIA test Bicyclo Prostaglandin E2 RIA test 11-Dehydrothromboxane B2 RIA test* 13,14-Dihydro-15-keto-PGF2a RIA test* 6-keto-PGF1a /2,3-dinor-6-keto-PGF1a RIA test Prostaglandin E2 RIA test Prostaglandin F2a RIA test* Thromboxane B2/2,3-dinor-TxB2 RIA test

 

 

Description - 500 test kit

The hTSH [125I] IRMA system provides a direct quantitative in vitro determination of human thyroid stimulating hormone (hTSH) in human serum. hTSH can be assayed in the range 0-100 µIU/ml using 100 or 200 µl serum sample. Each kit contains materials sufficient for 500 assay tubes.

Introduction

The thyroid stimulating hormone (thyrotropin or TSH) is a glycoprotein with a molecular weight of 28000, secreted by the adenohypophysis. Like other glycoprotein hormones (FSH, LH and HCG), TSH contains two different subunits, an alpha- and a ß-chain, linked by noncovalent bounds. The primary structure of alpha subunits of TSH and of the gonadotrophins is the same, whilst their ß subunits are different. The ß subunits are responsible for the immunological and biological specificity of these hormones.

The synthesis and the release of TSH are controlled by the circulatory level of thyroid hormones; triiodothyronine (T3) and thyroxin (T4) and by the hypothalamic thyrotropin releasing hormone (TRH). Thyroid hormones regulate the secretion of TSH by a negative feed-back mechanism. An elevation of T3 or T4 will suppress, and their fall will, in turn, increase the level of TSH in serum. The increased concentration of TSH in the serum is the earliest and best indicator of primary hypothyroidism.

The determination of TSH by immunoassay methods plays a crucial role in the diagnosis of thyroid disorders and in the evaluation of the functional integrity of the hypothalamic-pituitary axis.

The outstanding sensitivity of the present hTSH IRMA system makes it particularly suitable for the measurement of subnormal hTSH levels, a key to both the diagnosis and treatment follow up of hyperthyroid patients.

Principle of method

The technology uses two high affinity monoclonal antibodies in an immunoradiometric assay (IRMA) system.

The 125I labelled signal-antibody binds to an epitope of the TSH molecule spatially different from that recognised by the biotin-capture-antibody. The two antibodies react simultaneously with the antigen present in standards or samples, which leads to the formation of a capture antibody - antigen - signal antibody complex, also referred to as a “sandwich”.

During a 1 hour incubation period immuno-complex is immobilized to the reactive surface of streptavidin-coated test tubes. Reaction mixture is then discarded, test tubes washed exhaustively, and the radioactivity is measured in a gamma counter.

The concentration of antigen is directly proportional to the radioactivity measured in test tubes. By constructing a calibration curve plotting binding values against a series of calibrators containing known amount of hTSH, the unknown concentration of hTSH in patient samples can determined.

Contents of the kit

1 bottle TRACER, ready to use
105 ml, containing < 4.5 MBq
125I-signal and capture antibody in buffer with red dye and 0.1% NaN3.
8x2 vials STANDARD, ready to use.
1.5 ml per vial, containing 0 (S0), 0.06 (S0.06), 0.15 (S0.15), 0.6 (S0.6), 2.5 (S2.5), 15 (S15), 50 (S50) and 100 (S100) µIU/ml hTSH (WHO 2nd IRP 80/558) in serum with 0.1% NaN3
2x2 vials CONTROL SERUM low (CI) and high (CII)
1.5 ml, containing 0.1% NaN3

The concentration of the control serum is specified in the quality certificate enclosed.
10 boxes COATED TUBE, ready to use.
10 x 50 reactive test tubes, 12x75 mm, packed in plastic boxes.
1 bottle WASH BUFFER CONCENTRATE
100 ml, containing 0.1% NaN3. See Preparation of reagents
1 pc Quality certificate
1 pc Pack leaflet

Materials, tools and equipment required

Test tube rack
Precision pipettes with disposable tips (100, 200 µl and 2 ml)

Distilled water
Vortex mixer
Shaker
Plastic foil
Adsorbent tissue
Gamma counter

Recommended tools and equipment

Repeating pipettes (e.g. Eppendorf)
Dispenser with 1-L reservoir (instead of the 2 ml pipette)

Specimen collection and storage

Serum samples can be prepared according to common procedures used routinely in clinical laboratory practice. Samples can be stored at 2-8 °C if the assay is carried out within 24 hours, otherwise aliquots should be prepared and stored deep frozen (-20 °C). Frozen samples should be thawed and thoroughly mixed before assaying. Repeated freezing and thawing should be avoided. Do not use lipemic, hemolyzed or turbid specimens.

Preparation of reagents, storage

Add the wash buffer concentrate (100 ml) to 5 L distilled water to obtain 5.1 L wash solution. Upon dilution store at 2-8 °C until expiration.

Store the rest of reagents between 2-8 °C after opening. At this temperature each reagent is stable until expiry date. The actual expiry date is given on the package label and in the quality certificate.

CAUTION! Equilibrate all reagents and serum samples to room temperature. Mix all reagents and samples thoroughly before use. Avoid excessive foaming.

The Way of Use

The assay can be used in different procedures. There are three options for running the assay. The procedures are labelled as “A”, “B” and “C”.

Working in accordance to “A” and “B” you need a good laboratory shaker. In case of OPTION C you need a water bath thermostat. The test tubes must be in contact with the water so air conditioned thermostat is not applicable.

Patient sample values and expected values are the same for all procedures.

OPTION “A”: The Basic Procedure

It is very economical on sample consumption. Only 100 µl sample volume is needed. The sensitivity attainable is 0.011 µIU/ml. When the kit has less then 3 weeks to its expiration standard 0.06 µIU/ml can be omitted. However if sample consumption is not critical we recommend you to work according to OPTION B. Shaking is needed.

Standard solution 50 µIU/ml can also be omitted if the curve-fitting algorithm of the gamma counter gives similar results with or without this point.

OPTION “B”: The 3rd Generation Method

It works in the same way as OPTION “A” except the sample volume, which is 200 µl. The sensitivity attainable is 0.005 µIU/ml. Shaking is needed.

OPTION “C”: The Water Bath Procedure

No shaking is applied during incubation. A good laboratory thermostat is important. The sensitivity attainable is 0.020 µIU/ml. When the kit has less then 3 weeks to its expiration standard 0.06 µIU/ml can be omitted. Use this method if you have problems with your shakers but you need the results quickly. Use 200 µl sample volume only!

OPTION A - The Basic Procedure

(For a quick guide, refer to Table 1.)

1 Equilibrate reagents & samples to room temperature before use.
2 Label coated tubes in duplicate for each standard, control serum & samples.
3 Homogenize all reagents & samples by gentle mixing to avoid foaming.
4 Homogenize all reagents & samples by gentle mixing to avoid foaming.
5 Pipette 200 µl of tracer into each tube. (Set aside 2 tubes for total counts.)
6 Fix the test tube rack firmly onto the shaker plate. Turn on the shaker and adjust an adequate speed so that liquid is constantly rotating or shaking in each tube.
7 Incubate tubes for 1 hour shaking at room temperature.
8 Add 2.0 ml of diluted wash buffer to each tube. Decant the supernatant from all tubes by the inversion of the rack. In the upside down position place the rack on an absorbent paper for 2 minutes.
9 Return the tube-rack to an upright position, and repeat Step 8 two more times.
10 Count each tube for at least 60 seconds in a gamma counter.
11 Calculate the hTSH concentrations.

Table 1. Assay Protocol, Pipetting Guide - for OPTION A.
All volumes are in microliters (µl).

Tube

Total

Standard

Sample

Control

Standard

 

100

   

Sample

   

100

 

Control

     

100

Tracer

(200)

200

200

200

Shake for 1 hour at room temperature.

Wash buffer

 

2000

2000

2000

Decant the fluid and blot on filter paper

Wash buffer

 

2000

2000

2000

Decant the fluid and blot on filter paper

Wash buffer

 

2000

2000

2000

Decant the fluid and blot on filter paper

Count radioactivity (60 sec/tube)

Calculate the results

Characterization of the assay for OPTION A

Typical standard curve for the Turbo TSH I-125 IRMA kit (Option A)
Figure 1.
OPTION - A. Typical standard curve corresponding to Table 2.
(Do not use to calculate sample values!)

Table 2. Typical assay data for OPTION A

 

CPM
1

CPM
2

CPM
mean

B/T%

Total

406194

404399

405297

-

S0 (NSB)

58

67

63

0.015

S0.06

265

297

281

0.069

S0.15

615

587

601

0.148

S0.6

2158

2218

2188

0.54

S2.5

8377

8490

8434

2.08

S15

48394

50214

49304

12.16

S50

146568

147342

146955

36.26

S100 (Bmax)

243184

240624

241904

59.68

CI

1144

1201

1173

0.31

CII

62189

63957

63073

19.4

Sensitivity

The analytical sensitivity or minimum detectable limit is calculated by the interpolation of the mean counts of zero standard plus 2 standard deviation from the standard curve. Determination was carried out using 20 replicates of zero standard response.

The value of analytical sensitivity is 0.011 µIU/ml measured using fresh tracer.

The functional sensitivity is a measure of the hTSH concentration that is significantly different from zero as determined by the inter-assay precision profile (22% CV).

The value of functional sensitivity is: 0.07 µIU/ml.

Hook effect

There is no high dose hook effect up to an hTSH concentration 500 µIU/ml.

Precision

The within-assay (intra-assay) precision was determined with 15 replicates within a single run using pooled human serum samples. CV values are summarized below:

Table 3/1.

intra-assay

mean (µIU/ml)

CV %

0.179

3.8

0.738

3.5

6.25

4.2

The between-assay (inter-assay) precision was determined using pooled human serum samples in independent assay runs. The number of measurements on a sample was a function of sample volume available. Three different operators took part in the investigation process and four different tracer batches were used at different ages of the reagents. Four different lots of coated tubes were used to determine the inter-assay precision profile. Results are presented below.

Table 3/2.

 

inter-assay

number of assay runs

mean (µIU/ml)

CV %

13

0.093

19.5

19

2.98

4.4

19

20.37

7.4

Linearity - dilution test

Individual human serum samples were diluted with the zero standard of the kit. The diluted samples were measured according to kit protocol.

Table 4.

sample
No.

dilution factor

expected µIU/ml

observed µIU/ml

recovery
%

1

1

 

27.46

 

1

2.00

13.74

13.85

100.9

1

4.02

6.82

7.38

108.2

1

8.05

3.41

3.67

107.4

2

1

 

13.07

 

2

2.02

6.48

6.68

103.0

2

4.06

3.22

3.33

103.2

2

8.14

1.60

1.55

96.5

3

1

 

56.18

 

3

2.00

28.02

27.82

99.3

3

4.02

13.97

13.78

98.6

3

8.02

7.01

7.57

108.1

4

1

 

1.12

 

4

2.02

0.556

0.581

104.4

4

4.07

0.276

0.302

109.5

4

8.18

0.137

0.155

113.0

Recovery - addition test

Individual human serum samples were spiked with known amount of an elevated stock sample. Recovery % is to be interpreted as = (observed-base)/added*100. The results are summarized below.

Table 5.

sample

base µIU/ml

added µIU/ml

expected µIU/ml

observed µIU/ml

recovery
%

1

8.28

17.18

25.46

25.00

97.3

2

7.50

18.60

26.09

26.65

103.0

3

8.65

18.44

27.08

26.21

95.3

4

9.11

16.29

25.40

24.57

94.9

5

5.79

17.87

23.66

23.92

101.5

OPTION B - The 3rd Generation Method

(For a quick guide, refer to Table 6.)

1 Equilibrate reagents & samples to room temperature before use.
2 Label coated tubes in duplicate for each standard, control serum & samples.
3 Homogenize all reagents & samples by gentle mixing to avoid foaming.
4 Pipette 200 µl of standards, control & samples into the properly labeled tubes. Use rack to hold the tubes. Do not touch or scratch the inner bottom of the tubes with pipette tip.
5 Pipette 200 µl of tracer into each tube. (Set aside 2 tubes for total counts.)
6 Fix the test tube rack firmly onto the shaker plate. Turn on the shaker and adjust an adequate speed so that liquid is constantly rotating or shaking in each tube.
7 Incubate tubes for 1 hour shaking at room temperature.
8 Add 2.0 ml of diluted wash buffer to each tube. Decant the supernatant from all tubes by the inversion of the rack. In the upside down position place the rack on an absorbent paper for 2 minutes.
9 Return the tube-rack to an upright position, and repeat Step 8 two more times.
10 Count each tube for at least 60 seconds in a gamma counter.
11 Calculate the hTSH concentrations.

Table 6. Assay Protocol, Pipetting Guide for OPTION B
All volumes are in microliters (µl).

Tube

Total

Standard

Sample

Control

Standard

 

200

   

Sample

   

200

 

Control serum

     

200

Tracer

(200)

200

200

200

Shake for 1 hour at room temperature.

Wash buffer

 

2000

2000

2000

Decant the fluid and blot on filter paper

Wash buffer

 

2000

2000

2000

Decant the fluid and blot on filter paper

Wash buffer

 

2000

2000

2000

Decant the fluid and blot on filter paper

Count radioactivity (60 sec/tube)

Calculate the results

Characterization of the assay for OPTION B

Typical standard curve for the Turbo TSH I-125 IRMA kit
Figure 2.
OPTION - B. Typical standard curve corresponding to Table 7.
(Do not use to calculate sample values!)

Table 7. Typical assay data for OPTION B

 

CPM
1

CPM
2

CPM
mean

B/T%

Total

405254

402194

403724

-

S0 (NSB)

57

49

53

0.013

S0.06

468

435

452

0.069

S0.15

1095

1035

1065

0.264

S0.6

3936

4151

4044

1.00

S2.5

16214

15233

15724

3.89

S15

86741

88695

87718

21.73

S50

232356

229654

231005

57.22

S100 (Bmax)

311395

312518

311957

77.27

CI

2256

2099

2177

0.33

CII

115658

118590

117124

19.9

Sensitivity

The analytical sensitivity or minimum detectable limit is calculated by the interpolation of the mean counts of zero standard plus 2 standard deviation from the standard curve. Determination was carried out using 20 replicates of zero standard response.

The value of analytical sensitivity is 0.005 µIU/ml measured using fresh tracer.

The functional sensitivity is a measure of the hTSH concentration that is significantly different from zero as determined by the inter-assay precision profile (22% CV).

The value of functional sensitivity is: 0.03 µIU/ml.

Hook effect

There is no high dose hook effect up to an hTSH concentration 500 µIU/ml.

Precision

The within-assay (intra-assay) precision was determined with 15 replicates within a single run using pooled human serum samples. CV values are summarized below:

Table 8/1.

intra-assay

mean (µIU/ml)

CV %

0.091

8.3

0.527

3.2

18.55

2.1

The between-assay (inter-assay) precision was determined using pooled human serum samples in independent assay runs. The number of measurements on a sample was a function of sample volume available. Three different operators took part in the investigation process and four different tracer batches were used at different ages of the reagents. Four different lots of coated tubes were used to determine the inter-assay precision profile. Results are presented below.

Table 8/2.

 

inter-assay

number of assay runs

mean (µIU/ml)

CV %

21

0.062

8.7

22

1.20

2.9

22

18.83

2.6

Linearity - dilution test

Individual human serum samples were diluted with the zero standard of the kit. The diluted samples were measured according to kit protocol.

Table 9.

sample
No.

dilution factor

expected µIU/ml

observed µIU/ml

recovery
%

1

1

 

27.73

 

1

2.00

13.87

13.71

98.8

1

4.02

6.89

7.35

106.6

1

8.05

3.45

3.39

98.4

2

1

 

13.07

 

2

2.02

6.48

6.69

103.3

2

4.06

3.22

3.31

102.8

2

8.14

1.60

1.54

96.1

3

1

 

57.14

 

3

2.00

28.49

26.71

93.7

3

4.02

14.21

13.81

97.2

3

8.02

7.13

7.24

101.6

4

1

 

1.16

 

4

2.02

0.573

0.536

93.6

4

4.07

0.284

0.259

91.2

4

8.18

0.141

0.126

89.2

Recovery - addition test

Individual human serum samples were spiked with known amount of an elevated stock sample. Recovery % is to be interpreted as = (observed-base)/added*100. The results are summarised below.

Table 10.

sample
No.

base µIU/ml

added µIU/ml

expected µIU/ml

observed µIU/ml

recovery
%

1

10.08

15.71

25.78

26.24

102.9

2

14.63

13.33

27.96

27.38

95.6

3

9.88

13.49

23.37

22.76

95.5

4

0.769

14.27

15.03

15.04

100.0

5

1.50

14.63

16.13

16.01

99.2

OPTION C - The Water Bath Procedure

(For a quick guide, refer to Table 11.)

1 Equilibrate reagents & samples to room temperature before use.
2 Label coated tubes in duplicate for each standard, control serum & samples.
3 Homogenize all reagents & samples by gentle mixing to avoid foaming.
4 Pipette 200 µl of standards, control & samples into the properly labeled tubes. Use rack to hold the tubes. Do not touch or scratch the inner bottom of the tubes with pipette tip.
5 Pipette 200 µl of tracer into each tube. (Set aside 2 tubes for total counts.)
6 Vortex mix all tubes gently.
7 Incubate tubes for 1 hour in warm water barh thermostat (36-38 °C).
8 Add 2.0 ml of diluted wash buffer to each tube. Decant the supernatant from all tubes by the inversion of the rack. In the upside down position place the rack on an absorbent paper for 2 minutes.
9 Return the tube-rack to an upright position, and repeat Step 8 two more times.
10 Count each tube for at least 60 seconds in a gamma counter.
11 Calculate the hTSH concentrations.

Table 11. Assay Protocol, Pipetting Guide for OPTION C
All volumes are in microliters (µl).

Tube

Total

Standard

Sample

Control

Standard

 

200

   

Sample

   

200

 

Control serum

     

200

Tracer

(200)

200

200

200

Vortex mix
Incubate for 1 hour in water bath ( 36-38 °C )

Wash buffer

 

2000

2000

2000

Decant the fluid and blot on filter paper

Wash buffer

 

2000

2000

2000

Decant the fluid and blot on filter paper

Wash buffer

 

2000

2000

2000

Decant the fluid and blot on filter paper

Count radioactivity (60 sec/tube)

Calculate the results

Characterization of the assay for OPTION C

Typical standard curve for the Turbo TSH I-125 IRMA kit
Figure 3.
OPTION C Typical standard curve corresponding to Table 12.
(Do not use to calculate sample values!)

Table 12. Typical assay data for OPTION C

 

CPM
1

CPM
2

CPM
mean

B/T%

Total

402020

398564

400292

-

S0 (NSB)

71

60

66

0.017

S0.06

197

168

183

0.046

S0.15

347

388

368

0.092

S0.6

1361

1305

1333

0.333

S2.5

5339

5120

5230

1.307

S15

27317

26655

26986

6.74

S50

70967

68173

69570

17.38

S100 (Bmax)

87101

85910

86506

21.61

CI

754

698

726

0.32

CII

34536

35398

34967

19.5

Sensitivity

The analytical sensitivity or minimum detectable limit is calculated by the interpolation of the mean counts of zero standard plus 2 standard deviation from the standard curve. Determination was carried out using 20 replicates of zero standard response.

The value of analytical sensitivity is 0.020 µIU/ml measured using fresh tracer.

The functional sensitivity is a measure of the hTSH concentration that is significantly different from zero as determined by the inter-assay precision profile (22% CV).

The value of functional sensitivity is: 0.11 µIU/ml.

Hook effect

There is no high dose hook effect up to an hTSH concentration 500 µIU/ml.

Precision

The within-assay precision was determined with 15 replicates within a single run, the between-assay precision was estimated in 15 independent runs carried out in duplicates. CV values are summarized below:

Table 13/1.

intra-assay

mean (µIU/ml)

CV %

0.170

7.3

1.81

4.4

6.27

3.2

The between-assay (inter-assay) precision was determined using pooled human serum samples in independent assay runs. The number of measurements on a sample was a function of sample volume available. Three different operators took part in the investigation process and four different tracer batches were used at different ages of the reagents. Four different lots of coated tubes were used to determine the inter-assay precision profile. Results are presented below.

Table 13/2.

 

inter-assay

number of assay runs

mean (µIU/ml)

CV %

19

0.177

12.6

20

1.72

3.8

19

19.60

5.9

Linearity - dilution test

Individual human serum samples were diluted with the zero standard of the kit. The diluted samples were measured according to kit protocol.

Table 14.

sample
No.

dilution factor

expected µIU/ml

observed µIU/ml

recovery
%

1

1

 

11.31

 

1

2.02

5.59

5.96

106.6

1

4.09

2.77

3.04

109.8

1

8.31

1.36

1.38

101.6

2

1

 

24.57

 

2

2.02

12.14

11.61

95.7

2

4.10

5.99

5.88

98.2

2

8.33

2.95

3.08

104.2

3

1

 

33.46

 

3

2.02

16.60

17.60

106.0

3

4.08

8.20

9.44

115.2

3

8.21

4.07

4.90

120.1

Option independent data

Specificity

No cross reactivity with hFSH and hTSH can be detected in normal physiological concentrations. 2 000 mIU/ml hCG gives an apparent 3.5 µIU/ml increase in hTSH concentration.

Expected Values

Expected euthyroid range is 0.27 µIU/ml - 3.75 µIU/ml.

It is recommended that each laboratory determine a reference range for euthyroids for its own patient population, since this may vary in different laboratories or regions.

Calculation of results

The calculation is illustrated using representative data. The assay data collected should be similar to those shown in Table 2 / 7 / 12.

Calculate the average count per minute (cpm) for each pair of assay tubes. Calculate the normalized percent binding for each standard, control & sample respectively by using the following equation:

  Standard [Control, Sample] (cpm)  
B / T (%) =  ————————————   x 100
  T (cpm)  

Using logarithmic graph paper plot B/T (%) for each standard versus the corresponding concentration of hTSH.

Determine the hTSH concentration of the controls & unknown samples by interpolation from the standard curve. Automated data processing systems are also applicable.

Additional information

Components from various lots of this product, or those of different manufacturers should not be mixed or interchanged.

Note for the washing step: Decantation is the most critical step of the assay procedure. Pay a special attention not to contaminate the outer surface of tubes, when turning the test tube-rack upside down. Even a small contamination may introduce a high, unidentified background resulting in a substantial overestimation of concentration. The error associated may become particularly high in the low range of concentration, which is of vital importance for the reliable determination of subnormal TSH values. On the same reason, regular checking of the instrument background is inevitable. This is particularly important when multi-channel counters are used. Make sure that background values and variation between individual channels are within the range of acceptance as specified in counter's service book.

Note for the shaking step: The rack must hold the tubes tight during shaking in order to shake all the tubes with the same speed and strength.

Precaution

Radioactivity

This product contains radioactive material. It is the responsibility of the user to ensure that local regulations or code of practice related to the handling of radioactive materials are satisfied.

Biohazard

Human blood products used in the kit have been obtained from healthy human donors. They were tested individually by using approved methods (EIA, enzyme immunoassay), and were found to be negative, for the presence of both Human Immunodeficiency Virus antibody (Anti-HIV-1) and Hepatitis B surface Antigen (HBsAg).

Care should always be taken when handling human specimens to be tested with diagnostic kits. Even if the subject has been tested, no method can offer complete assurance that Hepatitis B Virus, Human Immunodeficiency Virus (HIV-1), or other infectious agents are absent. Human blood samples should therefore be handled as potentially infectious materials.

Chemical hazard

Components contain sodium azide as an antimicrobial agent. Dispose of waste by flushing with copious amount of water to avoid build-up of explosive metallic azides in copper and lead plumbing. The total azide present in each pack is 74 mg.

 

Use by In vitro diagnostic medical device Control
Batch code Manufacturer Standard
Caution, consult accompanying documents Radioactive material Coated tube
Biological risk Temperature limitation
Store between 2-8 °C
Tracer
Consult operating instructions Catalogue number Wash buffer


  Frequently asked questions about TurboTSH
Other thyroid function assays    
Calcitonin IRMA test RK-83M    
Anti-hTG RIA test RK-8CT T4 RIA test RK-11M1
TG IRMA test RK-51CT T4 RIA test RK-11M5
Anti-TPO RIA test RK-36CT T4 RIA test RK-11CT1
Turbo TSH IRMA test RK-1M1 T4 RIA test RK-11CT5
Turbo TSH IRMA test RK-1M2 Free T3 RIA test RK-33CT
Turbo TSH IRMA test RK-1M5 T3 RIA test RK-6M1
Turbo TSH IRMA test RK-1CT1 T3 RIA test RK-6M5
Free T4 RIA test RK-34M1 T3 RIA test RK-6CT1
Free T4 RIA test RK-34CT1 T3 RIA test RK-6CT5
Other immunoassay kits      
    PRINTABLE VERSION
    PDF-FORMAT

Turbo tsh5